Skip to main content

Table 1 Parameters for a pair of point sources used in registering magnetic image relative to optical image

From: Scanning SQUID microscope system for geological samples: system integration and initial evaluation

Point source

 

X (mm)

Error (1σ)

Y (mm)

Error (1σ)

Z (mm)

Error (1σ)

M (Am2)

Error (1σ%)

Dec (°)

Error (1σ)

Inc (°)

Error (1σ)

A

NRM

3.803

0.002

2.694

0.003

−0.266

0.003

2.04E−10

2.8

171.5

0.7

−2.1

0.7

AFD 10 mT

2.370

0.001

2.555

0.001

−0.264

0.001

2.13E−10

0.9

171.3

0.2

0.2

0.2

AFD 20 mT

2.756

0.002

2.638

0.002

−0.213

0.003

2.03E−10

1.0

171.0

0.3

0.4

0.3

ARM (+X)

2.519

0.001

2.259

0.001

−0.275

0.001

2.16E−10

0.8

171.2

0.2

0.1

0.2

ARM (+Y)

2.737

0.002

2.035

0.003

−0.280

0.003

2.27E−10

2.4

170.6

0.6

0.3

0.6

ARM (−Z)

2.669

0.002

1.839

0.002

−0.258

0.002

2.20E−10

1.9

168.6

0.4

1.1

0.5

Average

2.809

 

2.336

 

−0.259

 

2.14E−10

     

STD

0.509

 

0.349

 

0.024

 

9.22E−12

     

B

NRM

63.702

0.002

27.048

0.002

−0.757

0.003

2.50E−10

1.2

97.4

0.2

−2.3

2.0

AFD 10 mT

62.20

0.005

26.831

0.004

−0.881

0.007

2.65E−10

1.9

99.1

0.4

−2.1

0.3

AFD 20 mT

62.678

0.005

26.754

0.005

−1.032

0.009

2.75E−10

2.2

99.6

0.4

−3.4

0.3

ARM (+X)

62.345

0.007

26.480

0.006

−1.204

0.012

3.08E−10

2.7

99.8

0.5

−6.0

0.4

ARM (+Y)

62.476

0.003

26.209

0.002

−0.831

0.004

2.57E−10

1.3

97.4

0.2

−3.0

0.2

ARM (−Z)

62.657

0.003

26.189

0.004

−0.772

0.004

2.48E−10

1.3

97.1

0.3

−2.9

0.2

Average

62.677

 

26.585

 

−0.913

 

2.67E−10

     

STD

0.534

 

0.350

 

0.174

 

2.24E−11

     

Difference

NRM

59.899

 

24.354

 

−0.491

       

AFD 10 mT

59.833

 

24.276

 

−0.617

       

AFD 20 mT

59.922

 

24.116

 

−0.819

       

ARM (+X)

59.826

 

24.221

 

−0.929

       

ARM (+Y)

59.739

 

24.174

 

−0.552

       

ARM (−Z)

59.988

 

24.350

 

−0.514

       

Average

59.868

 

24.249

 

−0.654

       

STD

0.087

 

0.096

 

0.179

       
  1. Error means the error expected for the position optimized by magnetic image assuming a magnetic dipole
  2. A lower-left point source, B upper-right point source, Difference subtracted A from B
  3. STD standard deviation, M intensity of magnetic moment
  4. dec declination (clockwise from +x direction), inc inclination (positive for lower hemisphere)