Skip to main content

Table 1 HPM performances

From: First in-orbit results of the vector magnetic field measurement of the High Precision Magnetometer onboard the China Seismo-Electromagnetic Satellite

Sensors

Content

Performances

FGM

Measurement range

± 70000 nT

Linearity

Better than 0.005%

Bandwidth

15 Hz

Noise

Lower than 18 \( {\text{pT}}/\sqrt {\text{Hz}} \) @1 Hz

Stability

Better than 1 nT/month

Offset thermal drift

0.05 nT/ °C (before correction)

Gain thermal drift

5 ppm/ °C (before correction)

CDSM

Measurement range

1000–100,000 nT

Accuracy

0.19 nT (1σ)

Noise

Lower than 50 pT @1 s integration time

  1. Some non-linear correction parameters are considered during the ground calibration test