Table 2 Parameters of 20-cm and 5-cm SC sensors

From: Development of chopper-stabilized ASIC preamplifier for improving noise equivalent magnetic induction of search coil magnetometer probing space plasma waves

Parameter

20-cm SC

5-cm SC

Unit

Resistance of coil \(R\)

1.39

1.48

\({\text{k}}\Omega\)

Inductance of coil \(L\)

20.1

2.7

H

Parasitic capacitance of coil and 5-m harness \(C\)

1.00

nF

Number of windings of coil \(N\)

14,000

10,000

turns

Cross-sectional area of magnetic core \(S\)

5 × 5

mm2

Effective permeability of magnetic core \(\mu_{{{\text{eff}}}}\)

452

50