Skip to main content

Table 7 Mean values of the calibration parameters for each platform magnetometer dataset. The time-series are shown in Fig. 5

From: Co-estimating geomagnetic field and calibration parameters: modeling Earth’s magnetic field with platform magnetometer data

Dataset

\(b_1\)

\(b_2\)

\(b_3\)

\(s_1\)

\(s_2\)

\(s_3\)

\(u_1\)

\(u_2\)

\(u_3\)

(eu)

(eu)

(eu)

(eu/nT)

(eu/nT)

(eu/nT)

(\({^{\circ }}\))

(\({^{\circ }}\))

(\({^{\circ }}\))

CryoSat-2 FGM1

5.0

165.6

− 10.7

1.005178

1.004851

1.004479

0.453

0.191

− 0.336

CryoSat-2 FGM2

77.6

− 16.6

61.8

1.004697

1.003993

1.003427

− 0.288

0.050

0.502

CryoSat-2 FGM3

-115.2

− 29.4

− 44.6

1.000863

1.005424

1.002168

0.745

− 0.045

− 0.000

GRACE-A

746.4

− 2632.1

− 2310.0

1.034238

1.032041

1.018168

− 0.251

− 0.161

0.048

GRACE-B

406.0

− 2622.0

− 2005.6

1.029785

1.026781

1.017845

− 0.056

− 0.209

0.106